Cytek Aurora CS Cell Sorter

Cytek Aurora CSThe Cytek Aurora uses full spectrum flow cytometry and detects entire fluorochrome emission spectrum. This enables the resolution of fluorophores that are highly overlapped and cannot be resolved by the conventional flow cytometry. Combinations like APC and Alexa Fluor 647 can be clearly resolved, even if the populations are co-expressed. The cell sorter has the following features.

  • Full Spectrum Profiling (FSP) technology that improves data clarity and resolution with autofluorescence extraction
  • Enclosed in class II biosafety cabinet
  • Five lasers: 355, 405, 488, 561 and 640 nm
  • 3 scattering and 64 fluorescence channels, unevenly spaced bandwidth up to 829 nm: UV detector module, 16 channels; violet detector module, 16 channels; blue detector module, 14 channels; yellow-green detector module, 10 channels; red detector module, 8 channels
  • High sensitivity Coarse Wavelength Division Multiplexing (CWDM) semiconductor array enabling more efficient spectrum capture in the 365-829 nm range 
  • Sample flow rates adjustable in increments of 7 µL/min from 10 µL/min to 80 µL/min
  • Sample input formats: 12 x 75 mm or 15 mL tube with sample mixing
  • Sort output formats: 2- way sorting in 15 mL tube; up to 6-way sorting in 5 mL or 1.5 mL tubes; 96-well and 384-well plate with index sorting; custom plate options
  • Temperature control: 4 to 37 oC for both sample input and output
  • Spectroflo CS software for live unmixing during acquisition and sorting, autofluorescence extraction

Reserve the tool